Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Soar 3210: (1 year) CAL VER /5520 DATE: 10-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 107 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DC VOLTAGE} 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.007 DISP [32] 5520A NORMAL LO to UUT COM 1.008 DISP Turn the FUNCTION switch to DC V. 1.009 HEAD DC VOLTAGE: {300mV Range} 1.010 5520 300 300.0mV 0.6% 0.2U 2W 2.001 HEAD DC VOLTAGE: {3V Range} 2.002 5520 3 3.000V 0.8% 0.001U 2W 3.001 HEAD DC VOLTAGE: {30V Range} 3.002 5520 30 30.00V 0.8% 0.01U 2W 4.001 HEAD DC VOLTAGE: {300V Range} 4.002 5520 300 300.0V 0.8% 0.1U 2W 5.001 HEAD DC VOLTAGE: {1000V Range} 5.002 5520 1000 1000V 0.8% 1U 2W 6.001 HEAD {AC VOLTAGE} 6.002 DISP Turn the FUNCTION switch to AC V. 6.003 HEAD AC VOLTAGE: {3V Range} 6.004 5520 3 3.000V 1.5% 0.004U 100H SI 2W 7.001 HEAD AC VOLTAGE: {30V Range} 7.002 5520 30 30.00V 1.5% 0.04U 100H SI 2W 8.001 HEAD AC VOLTAGE: {300V Range} 8.002 5520 300 300.0V 1.5% 0.4U 100H SI 2W 9.001 HEAD AC VOLTAGE: {750V Range} 9.002 5520 750 700V 1.5% 4U 100H SI 2W 10.001 HEAD {RESISTANCE} 10.002 DISP Turn the FUNCTION switch to OHMS. 10.003 HEAD RESISTANCE: {30 MOhm Range} 10.004 5520 30 19.00MZ 3% 0.01U 2W 11.001 HEAD RESISTANCE: {3 MOhm Range} 11.002 5520 3 1.900MZ 1.5% 0.001U 2W 12.001 HEAD RESISTANCE: {300 kOhm Range} 12.002 5520 300 190.0kZ 1% 0.1U 2W 13.001 HEAD RESISTANCE: {30 kOhm Range} 13.002 DISP Connect the 5520A to the UUT as follows: 13.002 DISP [32] 5520A AUX HI to UUT V/OHM 13.002 DISP [32] 5520A AUX LO to UUT COM 13.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 13.002 DISP [32] 5520A NORMAL LO to UUT COM 13.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 13.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 13.003 5520 30 19.00kZ 1% 0.01U CW 14.001 HEAD RESISTANCE: {3 kOhm Range} 14.002 5520 3 1.900kZ 1% 0.001U CW 15.001 HEAD RESISTANCE: {300 Ohm Range} 15.002 5520 300 190.0Z 1% 0.2U CW 16.001 HEAD {CONTINUITY} 16.002 DISP Press the MODE SELECTION button to select CONTINUITY. 16.003 5520 10Z S CW 16.004 EVAL Is there an AUDIBLE TONE present (continuity test)? 17.001 HEAD {DIODE} 17.002 PIC 3210v 17.003 DISP Press the MODE SELECTION button to select DIODE. 17.004 M5520 RNGLK 33V 17.005 5520 0.500V 0.3% 0.003U 2W 18.001 M5520 * 18.002 HEAD {DC CURRENT} 18.003 DISP Turn the FUNCTION switch to 10A. 18.004 DISP Connect the 5520A to the UUT as follows: 18.004 DISP [32] 5520A 20A to UUT A 18.004 DISP [32] 5520A AUX LO to UUT COM 18.005 HEAD DC CURRENT: {10A Range} 18.006 5520 10 5.00A 1.5% 0.02U HC 2W 19.001 HEAD {AC CURRENT} 19.002 DISP Press the MODE SELECTION button to select AC amps. 19.003 HEAD AC CURRENT: {10A Range} 19.004 5520 10 5.00A 2% 0.05U 100H SI HC 2W 20.001 END